# Microelectrode array

> Mediated Wiki article. Canonical URL: https://mediated.wiki/source/Microelectrode_array
> Markdown URL: https://mediated.wiki/source/Microelectrode_array.md
> Source: https://en.wikipedia.org/wiki/Microelectrode_array
> Source revision: 1322732721
> License: Creative Commons Attribution-ShareAlike 4.0 International (https://creativecommons.org/licenses/by-sa/4.0/)

Device in neurophysiology

**Microelectrode arrays** (**MEAs**) (also referred to as multielectrode arrays[1][2]) are devices that contain multiple (tens to thousands) [microelectrodes](/source/Microelectrode) through which neural [signals](/source/Signal_(electronics)) are obtained or delivered, essentially serving as neural interfaces that connect [neurons](/source/Neuron) to [electronic circuitry](/source/Electric_circuit). There are two general classes of MEAs: implantable MEAs, used *[in vivo](/source/In_vivo)*, and non-implantable MEAs, used *[in vitro](/source/In_vitro)*. In each class, there are rigid, flexible, and [stretchable microelectrode array](/source/Stretchable_microelectrode_array).

## Theory

Neurons and [muscle](/source/Muscle) cells create [ion](/source/Ion) currents through their [membranes](/source/Cell_membrane) when excited, causing a change in [voltage](/source/Voltage) between the inside and the outside of the cell. When recording, the [electrodes](/source/Electrode) on an MEA [transduce](/source/Transducer) the change in [voltage](/source/Voltage) from the environment carried by [ions](/source/Ions) into currents carried by [electrons](/source/Electrons) (electronic currents). When stimulating, [electrodes](/source/Electrodes) transduce electronic currents into ionic currents through the media. This triggers the [voltage-gated ion channels](/source/Voltage-gated_ion_channel) on the [membranes](/source/Cell_membrane) of the excitable cells, causing the cell to [depolarize](/source/Depolarization) and trigger an [action potential](/source/Action_potential) if it is a neuron or a twitch if it is a muscle cell.[*[citation needed](https://en.wikipedia.org/wiki/Wikipedia:Citation_needed)*]

The size and shape of a recorded signal depend upon several factors: the nature of the medium in which the cell or cells are located (e.g. the medium's [electrical conductivity](/source/Electrical_conductivity), [capacitance](/source/Capacitance), and [homogeneity](/source/Homogeneous_(chemistry))); the nature of contact between the cells and the MEA electrode (e.g. area of contact and tightness); the nature of the MEA electrode itself (e.g. its geometry, [impedance](/source/Electrical_impedance), and noise); the [analog signal processing](/source/Analog_signal_processing) (e.g. the system's [gain](/source/Gain_(electronics)), [bandwidth](/source/Bandwidth_(signal_processing)), and behavior outside of [cutoff frequencies](/source/Cutoff_frequency)); and the data [sampling](/source/Sampling_(signal_processing)) properties (e.g. [sampling rate](/source/Sampling_rate) and [digital signal processing](/source/Digital_signal_processing)).[3] For the recording of a single cell that partially covers a planar electrode, the voltage at the [contact pad](/source/Contact_pad) is approximately equal to the voltage of the overlapping region of the cell and electrode multiplied by the ratio the [surface area](/source/Surface_area) of the overlapping region to the area of the entire electrode, or:

V p a d = V o v e r l a p × A o v e r l a p A e l e c t r o d e {\displaystyle V_{pad}=V_{overlap}\times {\frac {A_{overlap}}{A_{electrode}}}}

assuming the area around an electrode is [well-insulated](/source/Insulator_(electrical)) and has a very small capacitance associated with it.[3] The equation above, however, relies on modeling the electrode, cells, and their surroundings as an equivalent [circuit diagram](/source/Circuit_diagram). An alternative means of predicting cell-electrode behavior is by modeling the system using a geometry-based [finite element analysis](/source/Finite_element_analysis) in an attempt to circumvent the limitations of oversimplifying the system in a lumped circuit element diagram.[4]

An MEA can be used to perform [electrophysiological](/source/Electrophysiological) experiments on tissue slices or dissociated cell cultures. With acute tissue slices, the connections between the cells within the tissue slices prior to extraction and plating are more or less preserved, while the intercellular connections in dissociated cultures are destroyed prior to plating. With dissociated neuronal cultures, the neurons spontaneously form [networks](/source/Biological_neural_network).[5]

It can be seen that the voltage [amplitude](/source/Amplitude) an electrode experiences is [inversely related](/source/Inversely_related) to the distance from which a cell depolarizes.[6] Thus, it may be necessary for the cells to be cultured or otherwise placed as close to the electrodes as possible. With tissue slices, a layer of electrically passive dead cells form around the site of incision due to [edema](/source/Edema).[7] A way to deal with this is by fabricating an MEA with three-dimensional electrodes fabricated by [masking](/source/Photomask) and [chemical etching](/source/Etching_(microfabrication)). These 3-D electrodes penetrate the dead cell layer of the slice tissue, decreasing the distance between live cells and the electrodes.[8] In dissociated cultures, proper adherence of the cells to the MEA substrate is important for getting robust signals.

## History

The first implantable arrays were microwire arrays developed in the 1950s.[9] The first experiment involving the use of an array of planar electrodes to record from cultured cells was conducted in 1972 by C.A. Thomas, Jr. and his colleagues.[6] The experimental setup used a 2 x 15 array of [gold](/source/Gold) electrodes plated with [platinum black](/source/Platinum_black), each spaced 100 μm apart from each other. [Myocytes](/source/Myocytes) harvested from [embryonic](/source/Embryo) chicks were dissociated and cultured onto the MEAs, and signals up to 1 mV high in amplitude were recorded.[10] MEAs were constructed and used to explore the electrophysiology of snail [ganglia](/source/Ganglia) independently by [Guenter Gross](https://en.wikipedia.org/w/index.php?title=Guenter_Gross&action=edit&redlink=1) and his colleagues at the Center for Network Neuroscience in 1977 without prior knowledge of Thomas and his colleagues' work.[6] In 1982, Gross observed spontaneous electrophysiological activity from dissociated [spinal cord](/source/Spinal_cord) neurons, and found that activity was very dependent on temperature. Below about 30˚C signal amplitudes decrease rapidly to relatively small value at [room temperature](/source/Room_temperature).[6]

Before the 1990s, significant [entry barriers](/source/Entry_barrier) existed for new laboratories that sought to conduct MEA research due to the custom MEA fabrication and software they had to develop.[5] However, with the advent of affordable computing power[3] and commercial MEA hardware and software,[5] many other laboratories were able to undertake research using MEAs.

## Types

Microelectrode arrays can be divided up into subcategories based on their potential use: *in vitro* and *in vivo* arrays.

### *In vitro* arrays

An *in vitro* MEA

The standard type of *in vitro* MEA comes in a pattern of 8 x 8 or 6 x 10 electrodes. Electrodes are typically composed of [indium tin oxide](/source/Indium_tin_oxide), [platinum black](/source/Platinum_black) or [titanium nitride](/source/Titanium_nitride) and have diameters between 10 and 30 μm. These arrays are normally used for single-cell cultures or acute brain slices.[3]

One challenge among *in vitro* MEAs has been imaging them with [microscopes](/source/Microscopes) that use high power lenses, requiring low [working distances](https://en.wikipedia.org/w/index.php?title=Working_distance&action=edit&redlink=1) on the order of micrometers. In order to avoid this problem, "thin"-MEAs have been created using cover slip glass. These arrays are approximately 180 μm allowing them to be used with high-power lenses.[3][11]

Another challenge among *in vitro* MEAs has been the rigidity of the glass substrate, which does not replicate the soft, flexible nature of biological tissues, thus impacting cellular behavior and experimental outcomes.[12][13] To address this limitation, flexible and [stretchable microelectrode arrays](/source/Stretchable_microelectrode_array) have been developed to better simulate the mechanical properties of living tissues.[13][14] Manufacturers of flexible and stretchable MEAs such as BioMedical Sustainable Elastic Electronic Devices and Flexcell International Corporation are advancing MEA technologies to enhance the relevance of *in vitro* research by providing a more physiologically accurate environment for cells.[15][16]

In another special design, 60 electrodes are split into 6 × 5 arrays separated by 500 μm. Electrodes within a group are separated by 30 um with diameters of 10 μm. Arrays such as this are used to examine local responses of neurons while also studying functional connectivity of organotypic slices.[3][17]

Spatial resolution is one of the key advantages of MEAs and allows signals sent over a long distance to be taken with higher precision when a high-density MEA is used. These arrays usually have a square grid pattern of 256 electrodes that cover an area of 2.8 by 2.8 mm.[3]

Increased spatial resolution is provided by CMOS-based high-density microelectrode arrays featuring thousands of electrodes along with integrated readout and stimulation circuits on compact chips of the size of a thumbnail.[18] Even the resolution of signals propagating along single axons has been demonstrated.[19]

In order to obtain quality signals electrodes and tissue must be in close contact with one another. The perforated MEA design applies negative [pressure](/source/Pressure) to openings in the substrate so that tissue slices can be positioned on the electrodes to enhance contact and recorded signals.[3]

A different approach to lower the electrode impedance is by modification of the interface material, for example by using [carbon nanotubes](/source/Carbon_nanotubes),[20][21] or by modification of the structure of the electrodes, with for example gold nanopillars[22] or nanocavities.[23]

While long-term extracellular in vitro recording of neural cells has been the main avenue of the CMOS MEA,[24] its capacity has recently been extended to high-sensitivity multi-electrode intracellular in vitro recording of neural cells,[25] complementing the long-term extracellular recording.

### *In vivo* arrays

Schematic of the "Utah" in vivo electrode array

The three major categories of implantable MEAs are microwire, [silicon](/source/Silicon)-based,[26] and flexible microelectrode arrays. Microwire MEAs are largely made of stainless [steel](/source/Steel) or [tungsten](/source/Tungsten) and they can be used to estimate the position of individual recorded neurons by triangulation. Silicon-based microelectrode arrays include two specific models: the Michigan and Utah arrays. Michigan arrays allow a higher density of sensors for implantation as well as a higher spatial resolution than microwire MEAs. They also allow signals to be obtained along the length of the shank, rather than just at the ends of the shanks. In contrast to Michigan arrays, Utah arrays are 3-D, consisting of 100 conductive silicon needles. However, in a Utah array, signals are only received from the tips of each electrode, which limits the amount of information that can be obtained at one time. Furthermore, Utah arrays are manufactured with set dimensions and parameters while the Michigan array allows for more design freedom. Flexible arrays, made with [polyimide](/source/Polyimide), [parylene](/source/Parylene), or [benzocyclobutene](/source/Benzocyclobutene), provide an advantage over rigid microelectrode arrays because they provide a closer mechanical match, as the [Young's modulus](/source/Young's_modulus) of silicon is much larger than that of brain tissue, contributing to shear-induced [inflammation](/source/Inflammation).[9][13][27]

## Data processing methods

The fundamental unit of communication of neurons is, electrically, at least, the action potential. This all-or-nothing phenomenon originates at the [axon hillock](/source/Axon_hillock),[28] resulting in a depolarization of the intracellular environment which propagates down the [axon](/source/Axon). This ion flux through the cellular membrane generates a sharp change in voltage in the extracellular environment, which is what the MEA electrodes ultimately detect. Thus, voltage spike counting and sorting is often used in research to characterize network activity. Spike train analysis, can also save processing time and computing memory compared to voltage measurements. Spike timestamps are identified as times where the voltage measured by an individual electrode exceeds a threshold (often defined by standard deviations from the mean of an inactive time period). These timestamps can be further processed to identify bursts(multiple spikes in close proximity). Further analysis of these trains can reveal spike organization and temporal patterns.[29]

## Capabilities

### Advantages

In general, the major strengths of *in vitro* arrays when compared to more traditional methods such as [patch clamping](/source/Patch_clamp) include:[30]

- Allowing the placement of multiple electrodes at once rather than individually

- The ability to set up controls within the same experimental setup (by using one electrode as a control and others as experimental). This is of particular interest in stimulation experiments.

- The ability to select different recordings sites within the array

- The ability to simultaneously receive data from multiple sites

- Recordings from intact retinae are of great interest because of the possibility of delivering real-time optical stimulation and, for instance, the possibility of reconstructing receptive fields.

Furthermore, *in vitro* arrays are non-invasive when compared to patch clamping because they do not require breaching of the cell membrane.

With respect to *in vivo* arrays however, the major advantage over patch clamping is the high spatial resolution. Implantable arrays allow signals to be obtained from individual neurons enabling information such as position or [velocity](/source/Velocity) of motor movement that can be used to control a [prosthetic](/source/Prosthetic) device. Large-scale, parallel recordings with tens of implanted electrodes are possible, at least in rodents, during animal behavior. This makes such extracellular recordings the method of choice to identify of neural circuits and to study their functions. Unambiguous identification of the recorded neuron using multi-electrode extracellular arrays, however, remains a problem to date.

### Disadvantages

*In vitro* MEAs are less suited for recording and stimulating single cells due to their low spatial resolution compared to patch clamp and [dynamic clamp](https://en.wikipedia.org/w/index.php?title=Dynamic_clamp&action=edit&redlink=1) systems. The complexity of signals an MEA electrode could effectively transmit to other cells is limited compared to the capabilities of dynamic clamps.

There are also several biological responses to implantation of a microelectrode array, particularly in regards to chronic implantation.[31] Most notable among these effects are neuronal cell loss, [glial scarring](/source/Glial_scarring), and a drop in the number of functioning electrodes.[32] The tissue response to implantation is dependent among many factors including size of the MEA shanks, distance between the shanks, MEA material composition, and time period of insertion. The tissue response is typically divided into short term and long term response. The short term response occurs within hours of implantation and begins with an increased population of [astrocytes](/source/Astrocytes) and [glial cells](/source/Glial_cells) surrounding the device. The recruited [microglia](/source/Microglia) then initiate inflammation and a process of [phagocytosis](/source/Phagocytosis) of the foreign material begins. Over time, the astrocytes and microglia recruited to the device begin to accumulate, forming a sheath surrounding the array that extends tens of micrometres around the device. This not only increases the space between electrode probes, but also insulates the electrodes and increases impedance measurements. Problems with chronic implantation of arrays have been a driving force in the research of these devices. One novel study examined the [neurodegenerative](/source/Neurodegenerative) effects of inflammation caused by chronic implantation.[33] [Immunohistochemical](/source/Immunohistochemical) markers showed a surprising presence of hyperphosphorylated tau, an indicator of [Alzheimer's disease](/source/Alzheimer's_disease), near the electrode recording site. The phagocytosis of electrode material also brings into question the issue of a biocompatibility response, which research suggests has been minor and becomes almost nonexistent after 12 weeks *in vivo*. Research into minimizing the negative effects of device insertion includes surface coating of the devices with proteins that encourage neuron attachment, such as [laminin](/source/Laminin), or drug [eluting](/source/Elution) substances.[34][13]

## Applications

### *In vitro*

The nature of dissociated [neuronal networks](/source/Neuronal_network) does not seem to change or diminish the character of its [pharmacological](/source/Pharmacological) response when compared to *in vivo* models, suggesting that MEAs can be used to study pharmacological effects on dissociated neuronal cultures in a more simple, controlled environment.[35] A number of pharmacological studies using MEAs on dissociated neuronal networks, e.g. studies with [ethanol](/source/Ethanol).[36] Interlaboratory validation has been conducted using MEAs.[37]

In addition, a substantial body of work on various biophysical aspects of network function was carried out by reducing phenomena usually studied at the behavioral level to the dissociated cortical network level. For example, the capacity of such networks to extract spatial[38] and temporal[39] features of various input signals, dynamics of synchronization,[40] sensitivity to [neuromodulation](/source/Neuromodulation_(biology))[41][42][43] and kinetics of learning using closed loop regimes.[44][45] Finally, combining MEA technology with [confocal microscopy](/source/Confocal_microscopy) allows for studying relationships between network activity and synaptic remodeling.[11]

MEAs have been used to interface neuronal networks with non-biological systems as a controller. For example, a neural-computer interface can be created using MEAs. Dissociated rat [cortical](/source/Cerebral_cortex) neurons were integrated into a closed stimulus-response feedback loop to control an animat in a virtual environment.[46] A [closed-loop](/source/Feedback) stimulus-response system has also been constructed using an MEA by Potter, Mandhavan, and DeMarse,[47] and by Mark Hammond, [Kevin Warwick](/source/Kevin_Warwick), and Ben Whalley in the [University of Reading](/source/University_of_Reading). About 300,000 dissociated rat neurons were plated on an MEA, which was connected to motors and [ultrasound](/source/Ultrasound) sensors on a robot, and was conditioned to avoid obstacles when sensed.[48] Along these lines, Shimon Marom and colleagues in the [Technion](/source/Technion_%E2%80%93_Israel_Institute_of_Technology) hooked dissociated neuronal networks growing on MEAs to a [Lego Mindstorms](/source/Lego_Mindstorms) robot; the visual field of the robot was classified by the network, and commands were delivered to the robot wheels such that it completely avoids bumping into obstacles.[38] This ["Braitenberg vehicle"](/source/Braitenberg_vehicles) was used to demonstrate the [indeterminacy](/source/Underdetermination) of reverse neuro-engineering showing that even in a simple setup with practically unlimited access to every piece of relevant information,[49] it was impossible to deduce with certainty the specific [neural coding](/source/Neural_coding) scheme that was used to drive the robots behavior.

MEAs have been used to observe network firing in [hippocampal](/source/Hippocampal) slices.[50]

### *In vivo*

There are several implantable interfaces that are currently available for consumer use including [deep brain stimulators](/source/Deep_brain_stimulation), [cochlear implants](/source/Cochlear_implants), and [cardiac pacemakers](/source/Artificial_pacemaker). Deep brain stimulation (DBS) has been effective at treating movement disorders such as [Parkinson's disease](/source/Parkinson's_disease),[51] and cochlear implants have helped many to improve their hearing by assisting stimulation of the [auditory nerve](/source/Auditory_nerve). Because of their remarkable potential, MEAs are a prominent area of neuroscience research. Research suggests that MEAs may provide insight into processes such as memory formation and perception and may also hold therapeutic value for conditions such as [epilepsy](/source/Epilepsy), [depression](/source/Major_depressive_disorder), and [obsessive-compulsive disorder](/source/Obsessive-compulsive_disorder) [*[citation needed](https://en.wikipedia.org/wiki/Wikipedia:Citation_needed)*]. Clinical trials using interface devices for restoring motor control after spinal cord injury or as treatment for [ALS](/source/Amyotrophic_lateral_sclerosis) have been initiated in a project entitled BrainGate (see video demo: [BrainGate](http://www.cyberkineticsinc.com/video.htm)). MEAs provide the high resolution necessary to record time varying signals, giving them the ability to be used to both control and obtain feedback from prosthetic devices, as was shown by [Kevin Warwick](/source/Kevin_Warwick), [Mark Gasson](/source/Mark_Gasson) and [Peter Kyberd](/source/Peter_Kyberd).[52][53] Research suggests that MEA use may be able to assist in the restoration of vision by stimulating the [optic pathway](/source/Optic_nerve).[9]

## MEA user meetings

This section may contain information not important or relevant to the article's subject. Please help improve this section. (August 2020) (Learn how and when to remove this message)

A biannual scientific user meeting is held in [Reutlingen](/source/Reutlingen), organized by the Natural and Medical Sciences Institute (NMI) at the [University of Tübingen](/source/University_of_T%C3%BCbingen). The meetings offer a comprehensive overview of all aspects related to new developments and current applications of Microelectrode Arrays in basic and applied neuroscience as well as in industrial drug discovery, safety pharmacology and neurotechnology. The biannual conference has developed into an international venue for scientists developing and using MEAs from both industry and academia, and is recognized as an information-packed scientific forum of high quality. The meeting contributions are available as open access proceeding books.

## Use in art

In addition to being used for scientific purposes, MEAs have been used in [contemporary art](/source/Contemporary_art) to investigate philosophical questions about the relationship between technology and biology. Traditionally within Western thought, biology and technology have been separated into two distinct categories: *[bios](https://en.wiktionary.org/wiki/%CE%B2%CE%AF%CE%BF%CF%82#Ancient_Greek)* and *[technê.](/source/Techne)*[54] In 2002, *MEART: The Semi-living Artist* was created as a collaborative art and science project between [SymbioticA](/source/Symbiotica) at the [University of Western Australia](/source/University_of_Western_Australia) in [Perth](/source/Perth), and the Potter Lab at the [Georgia Institute of Technology](/source/Georgia_Institute_of_Technology) in [Atlanta](/source/Atlanta), to question the relationship between biology and technology.[55][56][57][58] *MEART* consisted of rat cortical neurons grown *in vitro* on an MEA in Atlanta, a pneumatic robot arm capable of drawing with pens on paper in Perth, and software to govern communications between the two. Signals from the neurons were relayed in a closed-loop between Perth and Atlanta as the MEA stimulated the pneumatic arm. *MEART* was first exhibited to the public in the exhibition *Biofeel* at the [Perth Institute of Contemporary Arts](/source/Perth_Institute_of_Contemporary_Arts) in 2002.[57][59]

## See also

- [Animat](/source/Animat)

- [Artificial cardiac pacemaker](/source/Artificial_cardiac_pacemaker)

- [Deep brain stimulation](/source/Deep_brain_stimulation)

- [Patch clamp](/source/Patch_clamp)

- [Bioelectronics](/source/Bioelectronics)

## References

1. **[^](#cite_ref-1)** The common misnomer multielectrode array replaces the prefix micro- with multi-, whose meaning is redundant with array. Micro- is important, as a small electrode is necessary to resolve the activity of single cells. Note that a microelectrode array is used for multichannel recordings.

1. **[^](#cite_ref-2)** ["What is a Multielectrode Array (MEA)?"](https://www.bmseed.com/multielectrode-array).

1. ^ [***a***](#cite_ref-Boven_3-0) [***b***](#cite_ref-Boven_3-1) [***c***](#cite_ref-Boven_3-2) [***d***](#cite_ref-Boven_3-3) [***e***](#cite_ref-Boven_3-4) [***f***](#cite_ref-Boven_3-5) [***g***](#cite_ref-Boven_3-6) [***h***](#cite_ref-Boven_3-7) Boven, K.-H.; Fejtl, M.; Möller, A.; Nisch, W.; Stett, A. (2006). "On Micro-Electrode Array Revival". In Baudry, M.; Taketani, M. (eds.). *Advances in Network Electrophysiology Using Multi-Electrode Arrays*. New York: Springer. pp. 24–37. [ISBN](/source/ISBN_(identifier)) [0-387-25857-4](https://en.wikipedia.org/wiki/Special:BookSources/0-387-25857-4).

1. **[^](#cite_ref-Buitenweg_4-0)** Buitenweg, J. R.; Rutten, W. L.; Marani, E. (2003). ["Geometry-based finite element modeling of the electrical contact between a cultured neuron and a microelectrode"](https://research.utwente.nl/en/publications/geometry-based-finite-element-modeling-of-the-electrical-contact-between-a-cultured-neuron-and-a-microelectrode(3fcb8a54-484f-46a0-aecc-07b37d485d42).html). *IEEE Trans Biomed Eng*. **50** (4): 501–509. [Bibcode](/source/Bibcode_(identifier)):[2003ITBE...50..501B](https://ui.adsabs.harvard.edu/abs/2003ITBE...50..501B). [doi](/source/Doi_(identifier)):[10.1109/TBME.2003.809486](https://doi.org/10.1109%2FTBME.2003.809486). [PMID](/source/PMID_(identifier)) [12723062](https://pubmed.ncbi.nlm.nih.gov/12723062). [S2CID](/source/S2CID_(identifier)) [15578217](https://api.semanticscholar.org/CorpusID:15578217).

1. ^ [***a***](#cite_ref-Potter2001_5-0) [***b***](#cite_ref-Potter2001_5-1) [***c***](#cite_ref-Potter2001_5-2) Potter, S. M. (2001). "Distributed processing in cultured neuronal networks". *Prog Brain Res*. Progress in Brain Research. Vol. 130. pp. 49–62. [doi](/source/Doi_(identifier)):[10.1016/S0079-6123(01)30005-5](https://doi.org/10.1016%2FS0079-6123%2801%2930005-5). [ISBN](/source/ISBN_(identifier)) [978-0-444-50110-3](https://en.wikipedia.org/wiki/Special:BookSources/978-0-444-50110-3). [PMID](/source/PMID_(identifier)) [11480288](https://pubmed.ncbi.nlm.nih.gov/11480288).

1. ^ [***a***](#cite_ref-PineBook_6-0) [***b***](#cite_ref-PineBook_6-1) [***c***](#cite_ref-PineBook_6-2) [***d***](#cite_ref-PineBook_6-3) Pine, J. (2006). "A History of MEA Development". In Baudry, M.; Taketani, M. (eds.). *Advances in Network Electrophysiology Using Multi-Electrode Arrays*. New York: Springer. pp. 3–23. [ISBN](/source/ISBN_(identifier)) [0-387-25857-4](https://en.wikipedia.org/wiki/Special:BookSources/0-387-25857-4).

1. **[^](#cite_ref-Heuschkel_7-0)** Heuschkel, M. O.; Wirth, C.; Steidl, E. M.; Buisson, B. (2006). "A History of MEA Development". In Baudry, M.; Taketani, M. (eds.). *Advances in Network Electrophysiology Using Multi-Electrode Arrays*. New York: Springer. pp. 69–111. [ISBN](/source/ISBN_(identifier)) [0-387-25857-4](https://en.wikipedia.org/wiki/Special:BookSources/0-387-25857-4).

1. **[^](#cite_ref-Thiebaud_8-0)** Thiebaud, P.; deRooij, N. F.; Koudelka-Hep, M.; Stoppini, L. (1997). "Microelectrode arrays for electrophysiological monitoring of hippocampal organotypic slice cultures". *IEEE Trans Biomed Eng*. **44** (11): 1159–63. [doi](/source/Doi_(identifier)):[10.1109/10.641344](https://doi.org/10.1109%2F10.641344). [PMID](/source/PMID_(identifier)) [9353996](https://pubmed.ncbi.nlm.nih.gov/9353996). [S2CID](/source/S2CID_(identifier)) [22179940](https://api.semanticscholar.org/CorpusID:22179940).

1. ^ [***a***](#cite_ref-Cheung_9-0) [***b***](#cite_ref-Cheung_9-1) [***c***](#cite_ref-Cheung_9-2) Cheung, K. C. (2007). "Implantable microscale neural interfaces". *Biomedical Microdevices*. **9** (6): 923–38. [doi](/source/Doi_(identifier)):[10.1007/s10544-006-9045-z](https://doi.org/10.1007%2Fs10544-006-9045-z). [PMID](/source/PMID_(identifier)) [17252207](https://pubmed.ncbi.nlm.nih.gov/17252207). [S2CID](/source/S2CID_(identifier)) [37347927](https://api.semanticscholar.org/CorpusID:37347927).

1. **[^](#cite_ref-Thomas_10-0)** Thomas, C. A.; Springer, P. A.; Loeb, G. E.; Berwald-Netter, Y.; Okun, L. M. (1972). "A miniature microelectrode array to monitor the bioelectric activity of cultured cells". *Exp. Cell Res*. **74** (1): 61–66. [doi](/source/Doi_(identifier)):[10.1016/0014-4827(72)90481-8](https://doi.org/10.1016%2F0014-4827%2872%2990481-8). [PMID](/source/PMID_(identifier)) [4672477](https://pubmed.ncbi.nlm.nih.gov/4672477).

1. ^ [***a***](#cite_ref-tenacity_11-0) [***b***](#cite_ref-tenacity_11-1) Minerbi, A.; Kahana, R.; Goldfeld, L.; Kaufman, M.; Marom, S.; Ziv, N. E. (2009). ["Long-term relationships between synaptic tenacity, synaptic remodeling, and network activity"](https://www.ncbi.nlm.nih.gov/pmc/articles/PMC2693930). *PLOS Biol*. **7** (6) e1000136. [doi](/source/Doi_(identifier)):[10.1371/journal.pbio.1000136](https://doi.org/10.1371%2Fjournal.pbio.1000136). [PMC](/source/PMC_(identifier)) [2693930](https://www.ncbi.nlm.nih.gov/pmc/articles/PMC2693930). [PMID](/source/PMID_(identifier)) [19554080](https://pubmed.ncbi.nlm.nih.gov/19554080).

1. **[^](#cite_ref-12)** Lacour, Stéphanie P.; Benmerah, Samia; Tarte, Edward; FitzGerald, James; Serra, Jordi; McMahon, Stephen; Fawcett, James; Graudejus, Oliver; Yu, Zhe; Morrison, Barclay (2010-10-01). ["Flexible and stretchable micro-electrodes for in vitro and in vivo neural interfaces"](https://link.springer.com/article/10.1007/s11517-010-0644-8). *Medical & Biological Engineering & Computing*. **48** (10): 945–954. [doi](/source/Doi_(identifier)):[10.1007/s11517-010-0644-8](https://doi.org/10.1007%2Fs11517-010-0644-8). [ISSN](/source/ISSN_(identifier)) [1741-0444](https://search.worldcat.org/issn/1741-0444). [PMID](/source/PMID_(identifier)) [20535574](https://pubmed.ncbi.nlm.nih.gov/20535574).

1. ^ [***a***](#cite_ref-:0_13-0) [***b***](#cite_ref-:0_13-1) [***c***](#cite_ref-:0_13-2) [***d***](#cite_ref-:0_13-3) Boufidis, Dimitris; Garg, Raghav; Angelopoulos, Eugenia; Cullen, D. Kacy; Vitale, Flavia (2025-02-21). ["Bio-inspired electronics: Soft, biohybrid, and "living" neural interfaces"](https://www.ncbi.nlm.nih.gov/pmc/articles/PMC11845577). *Nature Communications*. **16** (1): 1861. [Bibcode](/source/Bibcode_(identifier)):[2025NatCo..16.1861B](https://ui.adsabs.harvard.edu/abs/2025NatCo..16.1861B). [doi](/source/Doi_(identifier)):[10.1038/s41467-025-57016-0](https://doi.org/10.1038%2Fs41467-025-57016-0). [ISSN](/source/ISSN_(identifier)) [2041-1723](https://search.worldcat.org/issn/2041-1723). [PMC](/source/PMC_(identifier)) [11845577](https://www.ncbi.nlm.nih.gov/pmc/articles/PMC11845577). [PMID](/source/PMID_(identifier)) [39984447](https://pubmed.ncbi.nlm.nih.gov/39984447).

1. **[^](#cite_ref-14)** Graudejus, Oliver; Morrison, Barclay; Goletiani, Cezar; Yu, Zhe; Wagner, Sigurd (2012-02-08). ["Encapsulating Elastically Stretchable Neural Interfaces: Yield, Resolution, and Recording/Stimulation of Neural Activity"](https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3788117). *Advanced Functional Materials*. **22** (3): 640–651. [doi](/source/Doi_(identifier)):[10.1002/adfm.201102290](https://doi.org/10.1002%2Fadfm.201102290). [ISSN](/source/ISSN_(identifier)) [1616-301X](https://search.worldcat.org/issn/1616-301X). [PMC](/source/PMC_(identifier)) [3788117](https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3788117). [PMID](/source/PMID_(identifier)) [24093006](https://pubmed.ncbi.nlm.nih.gov/24093006).

1. **[^](#cite_ref-15)** ["Advance in vitro research with BMSEED's stretchable MEAs. ✓ adaptable to cell movement ✓ high-fidelity recordings ✓ innovative design for flexibility ➔ Redefine your research potential!"](https://www.bmseed.com/stretchable-multielectrodearrays-for-in-vitro-research). *BMSEED*. Retrieved 2025-10-02.

1. **[^](#cite_ref-16)** ["Tension Systems | Flexcell® International Corporation"](https://www.flexcellint.com/product-systems/tension). *www.flexcellint.com*. Retrieved 2024-11-12.

1. **[^](#cite_ref-17)** Segev, R.; Berry II, M. J. (2003). "Recording from all of the ganglion cells in the retina". *Soc Neurosci Abstr*. **264**: 11.

1. **[^](#cite_ref-A1_18-0)** Hierlemann, A.; Frey, U.; Hafizovic, S.; Heer, F. (2011). "Growing Cells atop Microelectronic Chips: Interfacing Electrogenic Cells in Vitro with CMOS-based Microelectrode Arrays". *Proceedings of the IEEE*. **99** (2): 252–284. [doi](/source/Doi_(identifier)):[10.1109/JPROC.2010.2066532](https://doi.org/10.1109%2FJPROC.2010.2066532). [S2CID](/source/S2CID_(identifier)) [2578216](https://api.semanticscholar.org/CorpusID:2578216).

1. **[^](#cite_ref-A2_19-0)** Bakkum, D. J.; Frey, U.; Radivojevic, M.; Russell, T. L.; Müller, J.; Fiscella, M.; Takahashi, H.; Hierlemann, A. (2013). ["Tracking axonal action potential propagation on a high-density microelectrode array across hundreds of sites"](https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5419423). *Nature Communications*. **4** 2181. [Bibcode](/source/Bibcode_(identifier)):[2013NatCo...4.2181B](https://ui.adsabs.harvard.edu/abs/2013NatCo...4.2181B). [doi](/source/Doi_(identifier)):[10.1038/ncomms3181](https://doi.org/10.1038%2Fncomms3181). [PMC](/source/PMC_(identifier)) [5419423](https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5419423). [PMID](/source/PMID_(identifier)) [23867868](https://pubmed.ncbi.nlm.nih.gov/23867868).

1. **[^](#cite_ref-20)** Yu, Z.; et al. (2007). "Vertically Aligned Carbon Nanofiber Arrays Record Electrophysiological Signals from Hippocampal Slices". *Nano Lett*. **7** (8): 2188–95. [Bibcode](/source/Bibcode_(identifier)):[2007NanoL...7.2188Y](https://ui.adsabs.harvard.edu/abs/2007NanoL...7.2188Y). [doi](/source/Doi_(identifier)):[10.1021/nl070291a](https://doi.org/10.1021%2Fnl070291a). [PMID](/source/PMID_(identifier)) [17604402](https://pubmed.ncbi.nlm.nih.gov/17604402).

1. **[^](#cite_ref-21)** Gabay, T.; et al. (2007). "Electro-chemical and biological properties of carbon nanotube based multi-electrode arrays". *Nanotechnology*. **18** (3) 035201. [Bibcode](/source/Bibcode_(identifier)):[2007Nanot..18c5201G](https://ui.adsabs.harvard.edu/abs/2007Nanot..18c5201G). [doi](/source/Doi_(identifier)):[10.1088/0957-4484/18/3/035201](https://doi.org/10.1088%2F0957-4484%2F18%2F3%2F035201). [PMID](/source/PMID_(identifier)) [19636111](https://pubmed.ncbi.nlm.nih.gov/19636111). [S2CID](/source/S2CID_(identifier)) [44491589](https://api.semanticscholar.org/CorpusID:44491589).

1. **[^](#cite_ref-22)** Brüggemann, D.; et al. (2011). "Nanostructured gold microelectrodes for extracellular recording from electrogenic cells". *Nanotechnology*. **22** (26) 265104. [Bibcode](/source/Bibcode_(identifier)):[2011Nanot..22z5104B](https://ui.adsabs.harvard.edu/abs/2011Nanot..22z5104B). [doi](/source/Doi_(identifier)):[10.1088/0957-4484/22/26/265104](https://doi.org/10.1088%2F0957-4484%2F22%2F26%2F265104). [PMID](/source/PMID_(identifier)) [21586820](https://pubmed.ncbi.nlm.nih.gov/21586820). [S2CID](/source/S2CID_(identifier)) [20738358](https://api.semanticscholar.org/CorpusID:20738358).

1. **[^](#cite_ref-23)** Hofmann, B.; et al. (2011). ["Nanocavity electrode array for recording from electrogenic cells"](http://juser.fz-juelich.de/search?p=id:%22PreJuSER-16546%22). *Lab Chip*. **11** (6): 1054–8. [doi](/source/Doi_(identifier)):[10.1039/C0LC00582G](https://doi.org/10.1039%2FC0LC00582G). [PMID](/source/PMID_(identifier)) [21286648](https://pubmed.ncbi.nlm.nih.gov/21286648).

1. **[^](#cite_ref-24)** Tsai, David; Sawyer, Daniel; Bradd, Adrian; Yuste, Rafael; Shepard, Kenneth (2017). ["A very large-scale microelectrode array for cellular-resolution electrophysiology"](https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5702607). *Nature Communications*. **8** (1) 1802. [Bibcode](/source/Bibcode_(identifier)):[2017NatCo...8.1802T](https://ui.adsabs.harvard.edu/abs/2017NatCo...8.1802T). [doi](/source/Doi_(identifier)):[10.1038/s41467-017-02009-x](https://doi.org/10.1038%2Fs41467-017-02009-x). [PMC](/source/PMC_(identifier)) [5702607](https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5702607). [PMID](/source/PMID_(identifier)) [29176752](https://pubmed.ncbi.nlm.nih.gov/29176752).

1. **[^](#cite_ref-25)** Wang, Jun; Jung, Woo-Bin; Gertner, Rona; Park, Hongkun; Ham, Donhee (2025). "Synaptic connectivity mapping among thousands of neurons via parallelized intracellular recording with a microhole electrode array". *Nature Biomedical Engineering*. **9** (7): 1144–1154. [doi](/source/Doi_(identifier)):[10.1038/s41551-025-01352-5](https://doi.org/10.1038%2Fs41551-025-01352-5). [PMID](/source/PMID_(identifier)) [39934437](https://pubmed.ncbi.nlm.nih.gov/39934437).

1. **[^](#cite_ref-26)** Bhandari, R.; Negi, S.; Solzbacher, F. (2010). "Wafer Scale Fabrication of Penetrating Neural Electrode Arrays". *Biomedical Microdevices*. **12** (5): 797–807. [doi](/source/Doi_(identifier)):[10.1007/s10544-010-9434-1](https://doi.org/10.1007%2Fs10544-010-9434-1). [PMID](/source/PMID_(identifier)) [20480240](https://pubmed.ncbi.nlm.nih.gov/20480240). [S2CID](/source/S2CID_(identifier)) [25288723](https://api.semanticscholar.org/CorpusID:25288723).

1. **[^](#cite_ref-27)** Chen, Ying; Zhang, Yingchao; Liang, Ziwei; Cao, Yu; Han, Zhiyuan; Feng, Xue (2020-02-04). ["Flexible inorganic bioelectronics"](https://doi.org/10.1038%2Fs41528-020-0065-1). *npj Flexible Electronics*. **4** (1) 2: 1–20. [doi](/source/Doi_(identifier)):[10.1038/s41528-020-0065-1](https://doi.org/10.1038%2Fs41528-020-0065-1). [ISSN](/source/ISSN_(identifier)) [2397-4621](https://search.worldcat.org/issn/2397-4621).

1. **[^](#cite_ref-Angelides_28-0)** Angelides, K. J.; Elmer, L. W.; Loftus, D.; Elson, E. (1988). ["Distribution and lateral mobility of voltage-dependent sodium channels in neurons"](https://www.ncbi.nlm.nih.gov/pmc/articles/PMC2115131). *J. Cell Biol*. **106** (6): 1911–25. [doi](/source/Doi_(identifier)):[10.1083/jcb.106.6.1911](https://doi.org/10.1083%2Fjcb.106.6.1911). [PMC](/source/PMC_(identifier)) [2115131](https://www.ncbi.nlm.nih.gov/pmc/articles/PMC2115131). [PMID](/source/PMID_(identifier)) [2454930](https://pubmed.ncbi.nlm.nih.gov/2454930).

1. **[^](#cite_ref-29)** Dastgheyb, Raha M.; Yoo, Seung-Wan; Haughey, Norman J. (2020). ["MEAnalyzer – a Spike Train Analysis Tool for Multi Electrode Arrays"](https://doi.org/10.1007%2Fs12021-019-09431-0). *Neuroinformatics*. **18** (1): 163–179. [doi](/source/Doi_(identifier)):[10.1007/s12021-019-09431-0](https://doi.org/10.1007%2Fs12021-019-09431-0). [PMID](/source/PMID_(identifier)) [31273627](https://pubmed.ncbi.nlm.nih.gov/31273627). [S2CID](/source/S2CID_(identifier)) [195795810](https://api.semanticscholar.org/CorpusID:195795810).

1. **[^](#cite_ref-30)** Whitson, J.; Kubota, D.; Shimono, K.; Jia, Y.; Taketani, M. (2006). "Multi-Electrode Arrays: Enhancing Traditional Methods and Enabling Network Physiology". In Baudry, M.; Taketani, M. (eds.). *Advances in Network Electrophysiology Using Multi-Electrode Arrays*. New York: Springer. pp. 38–68. [ISBN](/source/ISBN_(identifier)) [0-387-25857-4](https://en.wikipedia.org/wiki/Special:BookSources/0-387-25857-4).

1. **[^](#cite_ref-31)** Salatino, Joseph W.; Ludwig, Kip A.; Kozai, Takashi D. Y.; Purcell, Erin K. (2017-11-10). ["Glial responses to implanted electrodes in the brain"](https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6261524). *Nature Biomedical Engineering*. **1** (11): 862–877. [doi](/source/Doi_(identifier)):[10.1038/s41551-017-0154-1](https://doi.org/10.1038%2Fs41551-017-0154-1). [ISSN](/source/ISSN_(identifier)) [2157-846X](https://search.worldcat.org/issn/2157-846X). [PMC](/source/PMC_(identifier)) [6261524](https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6261524). [PMID](/source/PMID_(identifier)) [30505625](https://pubmed.ncbi.nlm.nih.gov/30505625).

1. **[^](#cite_ref-32)** Biran, R.; Martin, D. C.; Tresco, P. A. (2005). "Neuronal cell loss accompanies the brain tissue response to chronically implanted silicon microelectrode arrays". *Experimental Neurology*. **195** (1): 115–26. [doi](/source/Doi_(identifier)):[10.1016/j.expneurol.2005.04.020](https://doi.org/10.1016%2Fj.expneurol.2005.04.020). [PMID](/source/PMID_(identifier)) [16045910](https://pubmed.ncbi.nlm.nih.gov/16045910). [S2CID](/source/S2CID_(identifier)) [14077903](https://api.semanticscholar.org/CorpusID:14077903).

1. **[^](#cite_ref-33)** McConnell GC, Rees HD, Levey AI, Gross RG, Bellamkonda RV. 2008. Chronic electrodes induce a local, neurodegenerative state: Implications for chronic recording reliability. *Society for Neuroscience*, Washington, D.C[*[citation not found](https://en.wikipedia.org/wiki/Template:Harvard_citation_documentation#Wikilink_to_citation_does_not_work)*]

1. **[^](#cite_ref-34)** He, W.; McConnell, G. C.; Bellamkonda, R. V. (2006). "Nanoscale laminin coating modulates cortical scarring response around implanted silicon microelectrode arrays". *Journal of Neural Engineering*. **3** (4): 316–26. [Bibcode](/source/Bibcode_(identifier)):[2006JNEng...3..316H](https://ui.adsabs.harvard.edu/abs/2006JNEng...3..316H). [doi](/source/Doi_(identifier)):[10.1088/1741-2560/3/4/009](https://doi.org/10.1088%2F1741-2560%2F3%2F4%2F009). [PMID](/source/PMID_(identifier)) [17124336](https://pubmed.ncbi.nlm.nih.gov/17124336). [S2CID](/source/S2CID_(identifier)) [22732939](https://api.semanticscholar.org/CorpusID:22732939).

1. **[^](#cite_ref-Gopal_35-0)** Gopal, K. V.; Gross, G. W. (2006). "Emerging Histotypic Properties of Cultured Neuronal Networks". In Baudry, M.; Taketani, M. (eds.). *Advances in Network Electrophysiology Using Multi-Electrode Arrays*. New York: Springer. pp. 193–214. [ISBN](/source/ISBN_(identifier)) [0-387-25857-4](https://en.wikipedia.org/wiki/Special:BookSources/0-387-25857-4).

1. **[^](#cite_ref-Xia_36-0)** Xia, Y. & Gross, G. W. (2003). "Histotypic electrophysiological responses of cultured neuronal networks to ethanol". *Alcohol*. **30** (3): 167–74. [doi](/source/Doi_(identifier)):[10.1016/S0741-8329(03)00135-6](https://doi.org/10.1016%2FS0741-8329%2803%2900135-6). [PMID](/source/PMID_(identifier)) [13679110](https://pubmed.ncbi.nlm.nih.gov/13679110).

1. **[^](#cite_ref-37)** Novellino, A; Scelfo, B; Palosaari, T; Price, A; Sobanski, T; Shafer, T; Johnstone, A; Gross, G; Gramowski, A; Scroeder, O; Jügelt, K; Chiappalone, M; Benfenati, F; Martinoia, S; Tedesco, M; Defranchi, E; D'Angelo, P; Whelan, M (April 27, 2011). ["Development of micro-electrode array based tests for neurotoxicity: assessment of interlaboratory reproducibility with neuroactive chemicals"](https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3087164). *Front. Neuroeng*. **4** (4): 4. [doi](/source/Doi_(identifier)):[10.3389/fneng.2011.00004](https://doi.org/10.3389%2Ffneng.2011.00004). [PMC](/source/PMC_(identifier)) [3087164](https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3087164). [PMID](/source/PMID_(identifier)) [21562604](https://pubmed.ncbi.nlm.nih.gov/21562604).

1. ^ [***a***](#cite_ref-Order_38-0) [***b***](#cite_ref-Order_38-1) Shahaf, G.; Eytan, D.; Gal, A.; Kermany, E.; Lyakhov, V.; Zrenner, C.; Marom, S. (2008). ["Order-based representation in random networks of cortical neurons"](https://www.ncbi.nlm.nih.gov/pmc/articles/PMC2580731). *PLOS Comput. Biol*. **4** (11) e1000228. [Bibcode](/source/Bibcode_(identifier)):[2008PLSCB...4E0228S](https://ui.adsabs.harvard.edu/abs/2008PLSCB...4E0228S). [doi](/source/Doi_(identifier)):[10.1371/journal.pcbi.1000228](https://doi.org/10.1371%2Fjournal.pcbi.1000228). [PMC](/source/PMC_(identifier)) [2580731](https://www.ncbi.nlm.nih.gov/pmc/articles/PMC2580731). [PMID](/source/PMID_(identifier)) [19023409](https://pubmed.ncbi.nlm.nih.gov/19023409).

1. **[^](#cite_ref-39)** Eytan, D.; Brenner, N.; Marom, S. (2003). ["Selective adaptation in networks of cortical neurons"](https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6740578). *J. Neurosci*. **23** (28): 9349–9356. [doi](/source/Doi_(identifier)):[10.1523/JNEUROSCI.23-28-09349.2003](https://doi.org/10.1523%2FJNEUROSCI.23-28-09349.2003). [PMC](/source/PMC_(identifier)) [6740578](https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6740578). [PMID](/source/PMID_(identifier)) [14561862](https://pubmed.ncbi.nlm.nih.gov/14561862).

1. **[^](#cite_ref-40)** Eytan, D.; Marom, S. (2006). ["Dynamics and effective topology underlying synchronization in networks of cortical neurons"](https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6674346). *J. Neurosci*. **26** (33): 8465–8476. [doi](/source/Doi_(identifier)):[10.1523/JNEUROSCI.1627-06.2006](https://doi.org/10.1523%2FJNEUROSCI.1627-06.2006). [PMC](/source/PMC_(identifier)) [6674346](https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6674346). [PMID](/source/PMID_(identifier)) [16914671](https://pubmed.ncbi.nlm.nih.gov/16914671).

1. **[^](#cite_ref-41)** Eytan, D.; Minerbi, A.; Ziv, N. E.; Marom, S. (2004). "Dopamine-induced dispersion of correlations between action potentials in networks of cortical neurons". *J Neurophysiol*. **92** (3): 1817–1824. [doi](/source/Doi_(identifier)):[10.1152/jn.00202.2004](https://doi.org/10.1152%2Fjn.00202.2004). [PMID](/source/PMID_(identifier)) [15084641](https://pubmed.ncbi.nlm.nih.gov/15084641).

1. **[^](#cite_ref-42)** Tateno, T.; Jimbo, Y.; Robinson, H. P. (2005). "Spatio-temporal cholinergic modulation in cultured networks of rat cortical neurons: spontaneous activity". *Neuroscience*. **134** (2): 425–437. [doi](/source/Doi_(identifier)):[10.1016/j.neuroscience.2005.04.049](https://doi.org/10.1016%2Fj.neuroscience.2005.04.049). [PMID](/source/PMID_(identifier)) [15993003](https://pubmed.ncbi.nlm.nih.gov/15993003). [S2CID](/source/S2CID_(identifier)) [22745827](https://api.semanticscholar.org/CorpusID:22745827).

1. **[^](#cite_ref-43)** Tateno, T.; Jimbo, Y.; Robinson, H. P. (2005). "Spatio-temporal cholinergic modulation in cultured networks of rat cortical neurons: evoked activity". *Neuroscience*. **134** (2): 439–448. [doi](/source/Doi_(identifier)):[10.1016/j.neuroscience.2005.04.055](https://doi.org/10.1016%2Fj.neuroscience.2005.04.055). [PMID](/source/PMID_(identifier)) [15979809](https://pubmed.ncbi.nlm.nih.gov/15979809). [S2CID](/source/S2CID_(identifier)) [6922531](https://api.semanticscholar.org/CorpusID:6922531).

1. **[^](#cite_ref-44)** Shahaf, G.; Marom, S. (2001). ["Learning in networks of cortical neurons"](https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6762268). *J. Neurosci*. **21** (22): 8782–8788. [doi](/source/Doi_(identifier)):[10.1523/JNEUROSCI.21-22-08782.2001](https://doi.org/10.1523%2FJNEUROSCI.21-22-08782.2001). [PMC](/source/PMC_(identifier)) [6762268](https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6762268). [PMID](/source/PMID_(identifier)) [11698590](https://pubmed.ncbi.nlm.nih.gov/11698590).

1. **[^](#cite_ref-45)** Stegenga, J.; Le Feber, J.; Marani, E.; Rutten, W. L. (2009). ["The effect of learning on bursting"](https://research.utwente.nl/en/publications/the-effect-of-learning-on-bursting(be00a8b9-bb42-4e8f-b675-891b7f23e5d2).html). *IEEE Trans Biomed Eng*. **56** (4): 1220–1227. [Bibcode](/source/Bibcode_(identifier)):[2009ITBE...56.1220S](https://ui.adsabs.harvard.edu/abs/2009ITBE...56.1220S). [doi](/source/Doi_(identifier)):[10.1109/TBME.2008.2006856](https://doi.org/10.1109%2FTBME.2008.2006856). [PMID](/source/PMID_(identifier)) [19272893](https://pubmed.ncbi.nlm.nih.gov/19272893). [S2CID](/source/S2CID_(identifier)) [12379440](https://api.semanticscholar.org/CorpusID:12379440).

1. **[^](#cite_ref-Animat_46-0)** DeMarse, T. B.; Wagenaar, D. A.; Blau, A. W.; Potter, S. M. (2001). ["The Neurally Controlled Animat: Biological Brains Acting with Simulated Bodies"](https://www.ncbi.nlm.nih.gov/pmc/articles/PMC2440704). *Autonomous Robots*. **11** (3): 305–10. [doi](/source/Doi_(identifier)):[10.1023/A:1012407611130](https://doi.org/10.1023%2FA%3A1012407611130). [PMC](/source/PMC_(identifier)) [2440704](https://www.ncbi.nlm.nih.gov/pmc/articles/PMC2440704). [PMID](/source/PMID_(identifier)) [18584059](https://pubmed.ncbi.nlm.nih.gov/18584059).

1. **[^](#cite_ref-potterrobot_47-0)** Potter, S. M.; Madhavan, R.; DeMarse, T. B. (2003). ["Long-term bidirectional neuron interfaces for robotic control, and in vitro learning studies"](https://authors.library.caltech.edu/27336/). [*Proceedings of the 25th Annual International Conference of the IEEE Engineering in Medicine and Biology Society (IEEE Cat. No.03CH37439)*](https://resolver.caltech.edu/CaltechAUTHORS:20111020-154335843). pp. 3690–3693. [doi](/source/Doi_(identifier)):[10.1109/IEMBS.2003.1280959](https://doi.org/10.1109%2FIEMBS.2003.1280959). [ISBN](/source/ISBN_(identifier)) [0-7803-7789-3](https://en.wikipedia.org/wiki/Special:BookSources/0-7803-7789-3). [S2CID](/source/S2CID_(identifier)) [12213854](https://api.semanticscholar.org/CorpusID:12213854).

1. **[^](#cite_ref-Marks_48-0)** Marks, P. (2008). "Rise of the rat-brained robots". *New Scientist*. **199** (2669): 22–23. [doi](/source/Doi_(identifier)):[10.1016/S0262-4079(08)62062-X](https://doi.org/10.1016%2FS0262-4079%2808%2962062-X).

1. **[^](#cite_ref-49)** Marom, S.; Meir, R.; Braun, E.; Gal, A.; Kermany, E.; Eytan, D. (2009). ["On the precarious path of reverse neuro-engineering"](https://www.ncbi.nlm.nih.gov/pmc/articles/PMC2691154). *Front Comput Neurosci*. **3**: 5. [doi](/source/Doi_(identifier)):[10.3389/neuro.10.005.2009](https://doi.org/10.3389%2Fneuro.10.005.2009). [PMC](/source/PMC_(identifier)) [2691154](https://www.ncbi.nlm.nih.gov/pmc/articles/PMC2691154). [PMID](/source/PMID_(identifier)) [19503751](https://pubmed.ncbi.nlm.nih.gov/19503751).

1. **[^](#cite_ref-Colgin_50-0)** Colgin, L. L.; Kramar, E. A.; Gall, C. M.; Lynch, G. (2003). "Septal modulation of excitatory transmission in hippocampus". *J Neurophysiol*. **90** (4): 2358–2366. [doi](/source/Doi_(identifier)):[10.1152/jn.00262.2003](https://doi.org/10.1152%2Fjn.00262.2003). [PMID](/source/PMID_(identifier)) [12840078](https://pubmed.ncbi.nlm.nih.gov/12840078).

1. **[^](#cite_ref-51)** Breit, S.; Schulz, J. B.; Benabid, A. L. (2004). "Deep Brain Stimulation". *Cell and Tissue Research*. **318** (1): 275–288. [doi](/source/Doi_(identifier)):[10.1007/s00441-004-0936-0](https://doi.org/10.1007%2Fs00441-004-0936-0). [PMID](/source/PMID_(identifier)) [15322914](https://pubmed.ncbi.nlm.nih.gov/15322914). [S2CID](/source/S2CID_(identifier)) [25263765](https://api.semanticscholar.org/CorpusID:25263765).

1. **[^](#cite_ref-52)** Warwick, K.; Gasson, M.; Hutt, B.; Goodhew, I.; Kyberd, P.; Andrews, B.; Teddy, P.; Shad, A. (2003). "The Application of Implant Technology for Cybernetic Systems". *Archives of Neurology*. **60** (10): 1369–1373. [doi](/source/Doi_(identifier)):[10.1001/archneur.60.10.1369](https://doi.org/10.1001%2Farchneur.60.10.1369). [PMID](/source/PMID_(identifier)) [14568806](https://pubmed.ncbi.nlm.nih.gov/14568806).

1. **[^](#cite_ref-53)** Schwartz, A. B. (2004). "Cortical Neural Prosthetics". *Annual Review of Neuroscience*. **27**: 487–507. [doi](/source/Doi_(identifier)):[10.1146/annurev.neuro.27.070203.144233](https://doi.org/10.1146%2Fannurev.neuro.27.070203.144233). [PMID](/source/PMID_(identifier)) [15217341](https://pubmed.ncbi.nlm.nih.gov/15217341).

1. **[^](#cite_ref-54)** Thacker, Eugene (2010) "What is Biomedia?" in "Critical Terms for Media Studies" University of Chicago Press. Chicago and London, pp118-30

1. **[^](#cite_ref-55)** Bakkum DJ, Gamblen PM, Ben-Ary G, Chao ZC, Potter SM (2007). ["MEART: The Semi-Living Artist"](https://www.ncbi.nlm.nih.gov/pmc/articles/PMC2533587). *Frontiers in Neurorobotics*. **1**: 5. [doi](/source/Doi_(identifier)):[10.3389/neuro.12.005.2007](https://doi.org/10.3389%2Fneuro.12.005.2007). [PMC](/source/PMC_(identifier)) [2533587](https://www.ncbi.nlm.nih.gov/pmc/articles/PMC2533587). [PMID](/source/PMID_(identifier)) [18958276](https://pubmed.ncbi.nlm.nih.gov/18958276).

1. **[^](#cite_ref-56)** Bakkum, Douglas J.; Shkolnik, Alexander C.; Ben-Ary, Guy; Gamblen, Phil; DeMarse, Thomas B.; Potter, Steve M. (2004). "Removing Some 'A' from AI: Embodied Cultured Networks". *Embodied Artificial Intelligence*. Lecture Notes in Computer Science. Vol. 3139. pp. 130–45. [doi](/source/Doi_(identifier)):[10.1007/978-3-540-27833-7_10](https://doi.org/10.1007%2F978-3-540-27833-7_10). [ISBN](/source/ISBN_(identifier)) [978-3-540-22484-6](https://en.wikipedia.org/wiki/Special:BookSources/978-3-540-22484-6).

1. ^ [***a***](#cite_ref-Group_2002_pp.60-68_57-0) [***b***](#cite_ref-Group_2002_pp.60-68_57-1) SymbioticA research Group (2002) MEART – the semi living artist (AKA Fish & Chips) Stage 2 pp.60-68. in BEAP, Biennale of Electronic Art, 2002: The Exhibitions. Thomas, Paul, Ed., Pub. Curtin University. [ISBN](/source/ISBN_(identifier)) [1 74067 157 0](https://en.wikipedia.org/wiki/Special:BookSources/1_74067_157_0).

1. **[^](#cite_ref-58)** Ben-Ary, G, Zurr, I, Richards, M, Gamblen, P, Catts, O and Bunt, S (2001) "Fish and Chips, The current Status of the Research by the SymbioticA research group" in Takeover, wer macht die Kunst von morgen (who's doing the art of tomorrow) pp. 141-147 Springer Vien.

1. **[^](#cite_ref-59)** ["BioFeel: biology + art"](https://web.archive.org/web/20140811173950/http://www.pica.org.au/view/BioFeel%3A+biology+%2B+art/275/). Perth Institute for Contemporary Art. Archived from [the original](http://www.pica.org.au/view/BioFeel%3A+biology+%2B+art/275/) on 2014-08-11.

---
Adapted from the Wikipedia article [Microelectrode array](https://en.wikipedia.org/wiki/Microelectrode_array) by Wikipedia contributors ([contributor history](https://en.wikipedia.org/wiki/Microelectrode_array?action=history)). Available under [Creative Commons Attribution-ShareAlike 4.0 International](https://creativecommons.org/licenses/by-sa/4.0/). Changes may have been made.